A-A-54473
A pressure-type safety switch that is included in the cover latch assembly.
An electric brake circuit with manual actuating switch for shortening the deceleration of the rotating head after the timer reaches the “off” position and shuts off electric power to the motor.
A minimum (5) foot exposed length three-wire power cord that terminates with a three-blade male grounding plug that is suitable for use with receptacles shown in ANSI Standard C73.11 or shown in Configuration No. 5-15 of NEMA Standard WD6-88.
Shall be supplied with a manual(s) which contains complete instructions for the operation and maintenance of the centrifuge.
Workmanship. The centrifuge shall be free from defects which detract from its appearance or impair its serviceability.
Unit. Each (EA). One centrifuge, as specified, constitutes one unit.
Quality Assurance Provisions.
Responsibility for inspection. Unless otherwise specified in the contract or purchase order, the contractor is responsible for the performance of all inspection requirements as specified herein. Except as otherwise specified in the contact or purchase order, the contractor may use his own or any facilities suitable for the performance of the inspection requirements specified herein, unless disapproved by the Government. The Government reserves the right to perform any of the inspections set forth in the specification where such inspections are deemed necessary to assure supplies and services conform to prescribed requirements.
Records. Records of examinations and tests performed by or for the contractor shall be maintained by the contractor and made available to the Government upon the Government’s request, at any time, or from time to time, during the performance of the contract and for a period of three years after delivery of the supplies to which such records relate.
Inspection. Inspection, as used herein, is defined as both examination (such as visual or auditory investigation without the use of special laboratory appliances or procedures) and testing (determination by technical means of physical and chemical properties) of the item.